Pegasus P1800 Rack Power Supply

The Pegasus P1800 Range rack power supply is a general purpose DC High Voltage Power Supply suitable for laboratory and industrial use. Standard remote control is by analogue signals. Should other controls be preferred please discuss this with our technical staff.

Available in a standard 19-inch rack. Reduced weight and ease of serviceability is ensured, as the supplies are air insulated below 20kV.

The rack power supply can be provided with either a negative or positive polarity with an option of grid and filament outputs being available, please contact us for more information when requesting a quote or placing an order.

Informations complémentaires

Input Voltage

230VAC

Input Frequency

50Hz, 60Hz

Output Voltage

10kV, 20kV, 30kV, 40kV, 50kV, 70kV, 90kV

Output Power

300W, 600W, 900W

Polarity

Negative, Positive

Output Adjustable Range

0 to 100% Voltage

Voltage Ripple

0. 1%

Power Factor

0.95

Operating Temperature

0°C, 10°C, 15°C, 35°C

Operating Humidity

0 to 95% non-condensing

Cooling Method

Air

Control

Remote

Monitoring

Local and Remote

Protection

Arcing, Overload, Short Circuit

Protection

  • A trip occurs after excessive sparking or the occurrence of an error. The counter in the circuit limits the spark rate.
  • Current limit provides overload and short circuit protection.
  • Primary over-current protection is provided by the inverter circuits.

Control Interface

The Pegasus P1800 power supplies can be controlled by connecting an external control interface to the Harting connector or the D-Sub connector. Genvolt can also supply specially-designed control panels for customer requirements. Please see more information in the data sheet.

Safety

  • This power supply contains hazardous voltages and stored energy. Contact with the output may result in fatal injury. It should only be used and maintained by trained personnel.
  • The area where the power supply is to be used should be kept clean and dry.
  • Keep a safe distance from the output connector and any items connected to it.
  • Ensure that a secure connection is made between the Earth side of the load and the green and yellow Earth lead.

Demande de renseignements sur ce produit

Notre équipe d’experts peut vous aider à répondre à toutes les questions que vous vous posez sur ce produit. Remplissez le formulaire pour être rappelé ou appelez-nous pendant les heures de bureau.

Suivez-nous...

ASSISTANCE TECHNIQUE POUR LES ALIMENTATIONS À HAUTE TENSION

Les ingénieurs de Genvolt ont également offert un support technique personnalisé à des centaines de clients, qui représentent d’innombrables agences, départements de recherche et organisations mondiales pour de multiples produits dans de nombreuses applications. Les personnes qui travaillent avec nous savent qu’elles peuvent faire confiance à notre vision, à nos produits et à nos services, car nous construisons des partenariats solides.

Notre vision est de fournir aux clients des technologies de haute tension un niveau élevé d’assistance et une expertise de pointe afin de leur apporter la meilleure solution à leurs besoins.

Quelle que soit votre alimentation haute tension
nous pouvons vous aider !

Cela fait partie de notre éthique où "les clients font partie de notre activité et nous sommes dans le cadre de leurs activités".